Known Test Case Issues – Test Executive

The following test cases are available for execution in Test Executive, but issues with the Keysight Command Line Packet Generator (CLPG) FPGA may affect the test results or instrument functionality:

FPGA 6.20.1033 image 90CA 5-27-2016

Test Case Issue
1.1.10 The Data Generator FPGA image needs to be reloaded after running this test case because of a defect in the FPGA (see below).
2.3.1 Sometime the Data Generator does not correctly acknowledge the first TC0 frame before sending the NAC, in this case both data frames are replayed.

 

The following FPGA issues are still seen on very rare occasions, For these test cases just rebooting the the Data Generator twice often works to clear the issue.

Test Case Issue
2.5.1 Data Generator does not correctly replace an AFC with a NAC, sometimes the CLPG sends an AFC for all the sent frames or some of the sent frames. It should NAC all sent frames.
2.5.2 Sometimes the Data Generator does not send the correct AFCs with the correct credit values. The Data Generator needs to be reloaded after running this test case (see below).
2.5.4 Sometimes the Data Generator does not send the correct AFCs with the correct credit values. The Data Generator needs to be reloaded after running this test case (see below).

 

To reload the Data Generator FPGA:

  1. In the Setup -> Connection Setup tab change the link mode from Data Gen. + Analyzer to Analyzer, press apply
  2. Then change the link mode from Analyzer back to Data Gen. + Analyzer, press apply
  3. In the Analyzer Setup tab, press Sync Analyzer to Link Configuration, press OK